Ultra highly sensitive turbidity sensors have been launched. These sensors have a resolution of 0.0002FTU and their zero point noise is within ±0.0005, which allows accurate measurement in open sea and in deep sea where it used to be hardly possible to measure.
In this model, newly developed sensors are integrated in the usual data logger of Compact-series. Two types of models are now available for shallow water with fouling removal wiper (ATU6W-CMP) and for deep water with pressure-resistance of 6500m (ATU6-CMP).
Calibration is performed with formazine solution and their
concentration is shown in FTU unit.

Specifications

Instrument ATU6W-CMP ATU6-CMP
Parameters

Turbidity and battery voltage

Method Infrared backscattering
Measuring range 0 to 10FTU or 0 to 10, 10 to 100FTU
(10-100FTU:option)
Resolution 0.0002FTU (0.002FTU for 10 to 100FTU)
Accuracy ±0.002FTU or ±2% of measured value
Communication RS-232C
Capacity 179178 data
Memory 2M byte flash memory
A/D conversion 16 bit digital
Measuring mode Continuous, bust and real-time mode
Measuring interval 0.5, 1, 2, 5, 10, 15, 20, 30 seconds
Burst time 1 to 1440 minutes
Sample numbers 10, 15, 20, 30, 60, 120, 180, 240, 300, 600, 1200
Warming up time 10 seconds
Power/Capacity "D" lithium battery/14Ah "C" lithium battery/7Ah
Current consumption Measuring:50mA
Wake-up:10mA
Using wiper:450mA(Max)
Measuring:90mA
Wake-up:10mA
Materials Housing:titanium
Optical sensor:epoxy resin
Housing:titanium
Optical sensor:epoxy
Weight
(incl. battery)

1080g in air
560g in
water

800g in air
480g in water
Pressure-resistance 200m depth 6500m depth
Other Automatic clearance
with wiper
High pressure
resistant
Sensor structure is different between models with and without wiper.
If optional range is adopted, the resolution at 0 to 10FTU range remains unchanged.
This instrument is made on comparison calibration, using Alec standard as reference.