Sensor specifications (in common)
Parameter Depth Temperature Salinity (C)
Sensor type Semiconductor Thermistor Inductive Cell
Measuring Range 0 to 100 m
(ACTD-DF)
0 to 1000 m
(ACTD-CMP)
-5 to 40 °C 0 to 40
(0 to 60 mS/cm)
Resolution 0.01 m
(ACTD-DF)
0.02 m
(ACTD-CMP)
0.01 °C
(ACTD-DF)
0.001 °C
(ACTD-CMP)
0.01 mS/cm
(ACTD-DF)
0.001 mS/cm
(ACTD-CMP)
Accuracy 0.3 % FS ± 0.02 °C ± 0.03
± 0.02 mS/cm
 
Overview

ACTD-DF is a direct-reading version of the COMPACT-CTD Lite. It connects through 100 m of cable to the D-400F deck unit, which displays the data and stores them for later upload to a PC or laptop. In this configuration, the CTD-Lite can be used for spot measurements or profiling.

Four alkaline C-cells power both the D-400F deck unit and the instrument. The D-400F has 2MB of flash memory and a clear LCD display showing the current data set.
The display is back-lit and can be clearly read at night or in daylight. The splash proof touch panel buttons are easy to manipulate even with wet or gloved hands. This makes the D-400F unit good to work under field conditions.


Display specifications
Parameters Time, Depth, Temp, Salinity (C)
Memory 2 M-byte Flash memory,
180,000 data
Power 4xC Alkaline battery
(20h available)
Material Epoxy resin
Water resistance JIS 4 th grade
Dimensions W240 x D100 x H95 (mm)
Weight about 1.05 kg (incl. batteries)
Others Data transferable to PC

Logger specifications
Cable Kevlar reinforced 6.5 mm dia.
50 m extendable up to 200 m
Dimensions 42 dia. x L300 mm
Material Titanium grade 2
Weight in air 1 kg approx (excl. cable)
Weight in water 0.8 kg approx (excl. cable)
Water resistance 1000 m depth equivalent